Combi hand-held measuring device ProfiLine pH/Cond 3320 set 2
Order number: 206H.1
Part number: AC100015306
£ 2123.41*
*Prices without VAT and shipping costs£ 2548.09*
*incl. VAT, plus shipping costsPrice Breakdown
Measure with precision in any environment with the Combi Hand-Held Measuring Device Profiline Ph/Cond 3320 Set 2, engineered for laboratories where reliability matters. This robust instrument stands ready to deliver accurate readings across a broad pH and conductivity range, ensuring you never miss a critical measurement. Whether in the field or at the bench, this device has .
What Combi Hand-Held Measuring Device Profiline Ph/Cond Delivers
- Conductivity measurements are precise, accurate within ±0.5% of the measured value.
- Its conductivity measuring range spans 0 µS/cm to 1000 mS/cm, accommodating various samples.
- Automatic resolution allows for readings as fine as 0.001 µS/cm or as coarse as 1 mS/cm.
- Three different connections cater to diverse needs: DIN 19262 (pH), banana (temperature), and specific (8-pole) connectors.
- With a generous data storage capacity of 200 manual and 5000 automatic data sets, you can collect and store a wealth of information.
Convenience & Value
Combi Hand-Held Measuring Device Profiline Ph/Cond is listed under Combi hand-held measuring device with the pack size and title shown on this page.
Ready for High-Use Environments
Teams selecting Combi Hand-Held Measuring Device Profiline Ph/Cond typically match the Combi hand-held measuring device category and the container format to the task.
| Name | Value | |
|---|---|---|
| Port | : | USB-B (PC) |
| Type | : | pH/Cond 3320 set 2 |
| Sensor | : | SenTix ® 41 (pH) TetraCon ® 325 (EC) |
| Channels | : | 2 |
| IP rating | : | 67 |
| L x W x H | : | 180 x 80 x 55 mm |
| Connection | : | DIN 19262 (pH), banana, 2 x 4 mm (°C), specific (8-pole) (EC) |
| Parameters | : | pH redox conductivity salinity TDS temperature |
| Data memory | : | 200 manual and 5000 automatic data sets |
| pH accuracy | : | ±0.005 |
| Power supply | : | 4x battery Mignon/AA 1.5 V |
| TDS accuracy | : | ±0.5% of the measured value |
| pH resolution | : | 0,001 |
| TDS resolution | : | automatic (1 mg/l-0.1 g/l) |
| Voltage accuracy | : | ±1 mV |
| Salinity accuracy | : | ±0.2 (0-42) |
| Voltage resolution | : | 1 mV |
| pH measuring range | : | -2 to +19 |
| Salinity resolution | : | 0.1 |
| TDS measuring range | : | 0 mg/l to 200 g/l ppm |
| Temperature accuracy | : | ±0.1 °C |
| Conductivity accuracy | : | ±0.5% of the measured value |
| Temperature resolution | : | 0.1 °C |
| Conductivity resolution | : | automatic (0.001 µS/cm-1 mS/cm) |
| Voltage measuring range | : | -2500 to +2500 mV |
| Salinity measuring range | : | 0 to 70 ppt |
| Temperature measuring range | : | -5 to +105 °C |
| Conductivity measuring range | : | 0 µS/cm to 1000 mS/cm |
| Reference temperature conductivity | : | +20 / +25 °C |