Produced for stability across temperature-sensitive applications, the Laboratory Cooling Water Monitor Aquastop II falls under the category of monitoring instruments, specifically monitors and flow meters. This device is designed to monitor cooling water flow and temperature within laboratory settings, particularly for equipment running without constant supervision. It features a 2-line LED display for simultaneous viewing of both flow rate and temperature, as well as an LED indicator for switching status and alarm notifications. Users have the flexibility to set their own customizable alarm limits, with acoustic pre-alarms alerting them prior to final deactivation. The device comes equipped with a Schuko socket on its rear side, and includes a solenoid valve, flow sensor with an internal NTC sensor, and a hose connector (for hoses with an inner diameter of 10mm). Additional specifications include a rated power of 3500W, a fuse rating of
| Name | Value | |
|---|---|---|
| Fuse | : | 16 A mT (load)0.5 A mT (solenoid valve)0.2 A mT (control unit) |
| Type | : | AQUASTOP ® II |
| Weight | : | 1300 g |
| L x W x H | : | 205 x 175 x 78 mm |
| Rated power | : | 3500 W |
| Rated current | : | 15 A |
| Solenoid valve | : | connection to tap: 3 / 4 ″ IG connection to apparatus: 3 / 4 ″ AG or hose connector Ø 10 mm |
| Mains connection | : | 230 V, 50 Hz |
| Setting accuracy | : | ±1 l/h, ±1 °C |
| Flow measuring range | : | 2-99 l/h |
| Ext. temp. measuring range | : | 0 to +250 °C |
| Temperature measuring range, internal | : | 0 to +80 °C |
© 2023 Copyright. All Rights Reserved. Unit 19, Euroway House, Roydsdale Way, Bradford, BD4 6SE, UK
© 2023 Copyright. All Rights Reserved. Unit 19, Euroway House, Roydsdale Way, Bradford, BD4 6SE, UK